Calibration Data Measurements
            
                
                
                
                
            
            
                | Barometric pressure (corrected) | 
                PB
                 | 
                kPa | 
                ±.340 kPa. | 
            
            
                | Ambient temperature | 
                TA
                 | 
                °C | 
                ±.28 °C. | 
            
            
                | Air temperature into metering venturi | 
                ETI | 
                °C | 
                ±1.11 °C. | 
            
            
                | Pressure drop between the inlet and throat of metering venturi | 
                EDP | 
                kPa | 
                ±0.012 kPa. | 
            
            
                | Air flow | 
                QS
                 | 
                m3/min. | 
                ±0.5 percent of NIST value. | 
            
            
                | Air temperature at CVS pump inlet | 
                PTI | 
                °C | 
                ±1.11 °C. | 
            
            
                | Pressure depression at CVS pump inlet | 
                PPI | 
                kPa | 
                ±0.055 kPa. | 
            
            
                | Pressure head at CVS pump outlet | 
                PPO | 
                kPa | 
                ±0.055 kPa. | 
            
            
                | Air temperature at CVS pump outlet (optional) | 
                PTO | 
                °C | 
                ±1.11 °C. | 
            
            
                | Pump revolutions during test period | 
                N | 
                Revs | 
                ±1 Rev. | 
            
            
                | Elapsed time for test period | 
                t | 
                s | 
                ±0.5 s. |