               	               		35.61—Calibration of survey instruments.
               		
               	 	
               	 	
               	 	               	 	               	 	               	 	               	 	
               	 	
    
        
        (a) 
         A licensee shall calibrate the survey instruments used to show compliance with this part and  10 CFR Part 20 before first use, annually, and following a repair that affects the calibration. A licensee shall—
     
    
        
        (1) 
         Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;
     
    
        
        (2) 
         Calibrate two separated readings on each scale or decade that will be used to show compliance; and
     
    
        
        (3) 
         Conspicuously note on the instrument the date of calibration.
     
    
        
        (b) 
         A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
     
    
        
        (c) 
         A licensee shall retain a record of each survey instrument calibration in accordance with  § 35.2061.